Instructions
Intended Use
This device is intended for medical purpos-
es to illuminate body surfaces. It is used for
non-invasive visual examination of intact
skin.
This battery-operated product is designed
for external examination only within pro-
fessional healthcare facilities by medical
professionals.
Check the correct operation of the device
before use! Do not use the device if there
are visible signs of damage.
CAUTION: Do not look directly into the LED
light. Patients must close their eyes during
examinations.
In case of a serious incident with the use
of this device, notify DermLite immediately
and, if required by local regulations, your
national health authority.
CAUTION: Do not use the device in fire
or explosive risk area (e.g. oxygen-rich
environment).
WARNING: This product can expose you to
chemicals including methylene chloride and
hexavalent chromium, which are known to
the State of California to cause cancer or
reproductive toxicity. For more information
go to www.P65Warnings.ca.gov.
Electromagnetic Compatibility
This device complies with the EMC Emis-
sions and Immunity level requirements of
the standard IEC 60601-1-2:2014. The
emission characteristics of this equipment
make it suitable for use in professional
healthcare environment as well as residen-
tial environment (CISPR 11 Class B). This
equipment offers adequate protection to
radio communication service. In the rare
event of interference to the radio communi-
cation service, the user might need to take
mitigation measures, such as relocating or
re-orienting equipment.
WARNING: Use of this equipment adjacent
to or stacked with other equipment should
be avoided because it could result in im-
proper operation. If such use is necessary,
this equipment and the other equipment
should be observed to verify that they are
operating normally.
WARNING: Use of accessories other than
those provided by the manufacturer of
this equipment could result in increased
electromagnetic emissions or decreased
electromagnetic immunity of this equipment
and result in improper operation.
WARNING: Portable RF transmitters
should be used no closer than 30 cm (12
inches) to any part of the device. Other-
wise, degradation of the performance of
this equipment could result.
IMPORTANT: Before initial use, please
charge the internal lithium ion battery. See
"Battery and Charging" for details.
DermLite DL200HR is a smartphone-com-
patible pocket dermatoscope with 10x
magnification intended for polarized exam-
ination of skin lesions, with or without skin
contact.
IceCap®
To support your infection control measures,
snap one of the included disposable Ice-
Caps (IC) over the installed faceplate (FP).
To remove and dispose, simply pull it off.
Hold your device so that the LEDs are
aimed at the skin. Position the device
©5mm above the skin and push the Power
button (P) to turn on the device. Look
through the lens with your eye 25-50 mm
above the lens. Move the device closer
or farther away from the skin to obtain the
desired image focus. For dermoscopy with
skin contact, or for camera use, turn the
Dial (D) to extend the Spacer (S) and focus
the image.
Push the Mode button (M) to choose be-
tween two brightness levels.
To turn off your DermLite, push the Power
button (P); it turns off automatically after 3
minutes.
Removing the Spacer.
The Spacer (S) may be removed by turning
the Dial (D) counter-clockwise beyond a
noticeable resistance, then pulling it off the
device. To attach the Spacer, insert its 3
legs into the 3 holes until resistance is felt.
Place the device level on a flat surface and
evenly push down while simultaneously
rotating the Dial (D) clockwise.
Removing the Faceplate.
The Faceplate (FP) may simply be pulled
off the Spacer (S) by grabbing it at the
fingernail notches. To reattach, simply snap
it back in place.
Photography.
Your DermLite DL200 may be attached
to many cameras or smartphones using
DermLite MagnetiConnect® adapters (sold
separately). Prior to taking photos, wipe
the Faceplate (FP) clean and extend the
Spacer (S).
Battery and Charging
CAUTION: This device uses a special 3.7V
1800 mAh lithium ion battery, which can
only be purchased from DermLite or from
an authorized DermLite dealer. Do not un-
der any circumstances use another battery
other than the one designed for this unit.
This device is equipped with a four-level
Charge Indicator (CI). When the unit is
turned on and fully charged, all four LEDs
light up. When less than 75% of battery
life remains, three LEDs are illuminated. If
more than half the battery is drained, two
LEDs are on, while a single LED indicates
that less than 25% of battery capacity
remains. To charge your device, connect
the supplied USB-C to USB cable to the
Charging Port (CP) and any IEC 60950-
1 compliant (5V) USB port. The Charge
Indicator (CI) pulses to indicate charging,
with the number of pulsing LEDs corre-
sponding with charge level. When charging
is complete, all four indicator LEDs illumi-
nate. After years of use, you may wish to
re-place the battery, which is only available
directly from DermLite or an authorized
DermLite dealer.
To remove the battery (B), detach the snap-
on Cover (C): Pry it open, starting at the
small notch (N), using a screwdriver, then
work your way around the Cover to release
it completely. Disconnect the Battery Plug
(BP), insert a new battery and reconnect.
Reattach the Cover securely.
Troubleshooting
Please check www.dermlite.com for the
most current troubleshooting information. If
your device requires servicing, visit www.
dermlite.com/service or contact your local
DermLite dealer.
Care and Maintenance
CAUTION: No modification of this equip-
ment is allowed.
Your device is designed for trouble-free
operation. Repairs shall be made only by
qualified service personnel.
Cleaning
The exterior of your device, except the
optical parts, may be wiped clean with
isopropyl alcohol (70% vol.) prior to use
on a patient. The lens should be treated
as high-quality photographic equipment
and should be cleaned with standard lens
cleaning equipment and protected from
harmful chemicals. Do not use abrasive
material on any part of the equipment or
immerse the device in liquid. Do not auto-
clave.
Warranty: 10 years for parts and labor. The
battery is warranted for 1 year.
Disposal
This device contains electronics and a lith-
ium-ion battery that must be separated for
disposal and may not be disposed of with
general household waste. Please observe
local disposal regulations.
The DermLite DL200 set includes:
DL200 dermatoscope (21 LEDs, 25 mm
10x lens, spacer, faceplate with 10 mm
reticle), silicone cover, lanyard, microfiber
case, USB charging cable, 4 IceCaps.
Technical Description
Visit www.dermlite.com/technical or contact
your local DermLite dealer.