7
Faults
If you are unable to remedy faults with the aid of the table, or in the event of unexpected
operation or an incident, contact your specialist dealer to have the device repaired. To avoid
serious damage, do not continue using the device.
Fault
does not appear on the
therapy device's display.
No values for SpO
and PR are
2
displayed.
8
Maintenance
The module is designed to have a useful service life of 6 years.
If the module is used as intended in accordance with these instructions for use, it does not
require any maintenance.
If you identify faulty parts during the function check
contact your authorized dealer.
Cause
Contact problem or defective
module.
Contact problem or defective
SpO
sensor.
2
The SpO
sensor is poorly
2
positioned
Nail polish or contaminated
finger
prisma CHECKEN EN 41
Remedy
Remove and then reconnect the
module
(see 4.2,
p.35).
If the fault persists: Have the module
replaced.
Disconnect the sensor from the SpO
connection cable or the SpO
connection cable from the module and
reconnect.
If the fault persists:
Have the sensor replaced.
Check the position and fitting of the
SpO
sensor. Select another sensor
2
size if necessary.
Check finger. If necessary: Clean
finger.
(see "6 Function check",
prisma CHECK
p.36),
(see 4.1,
2
2
page 40),
EN
41