Function check
6.1 Intervals
6.2 Performing a function check
Result
40
EN
prisma CHECK
Carry out a function check at regular intervals:
•
After each hygienic preparation
•
After each repair
•
At least every 6 months
1. Inspect the housing for cracks, damage, and heavy soiling.
2. Connect the module up to the therapy device
Connecting the module up to the therapy device",
3. Check whether the
device display.
4. Connect the pulse oximetry sensor
pulse oximetry sensor",
5. Check whether the current oxygen saturation and pulse rate
values are shown on the display.
6. If the module does not work properly or shows signs of
damage: Contact your authorized dealer.
The function check is complete.