5
Hygienic preparation
5.1 General information
5.2 Intervals
5.3 Hygienic preparation of the module
Requirement
Part
Housing
Connection cable
Result
•
Wear suitable protective equipment for disinfection work.
•
Please refer to the instructions for use supplied with the
disinfectant used.
Interval
Weekly
As necessary, e.g., following infectious
diseases or severe soiling
Damage to the device caused by ingress of liquids!
Ingress of liquids may damage the therapy device, components,
and accessories.
⇒ Do not immerse the therapy device, components, or
accessories in liquids.
The module is disconnected from the therapy device
Removing the module from the therapy device",
1. Carry out hygienic preparation of the module as specified in
the following table:
Cleaning
Wipe with a damp cloth
using water or mild soap
2. Allow the module housing and in particular the electrical
interfaces to dry well.
The module is hygienically prepared.
Action
Clean the modules
Disinfect the modules
Disinfection
Wipe disinfection
(Recommendation: terralin
protect or perform advanced
Alcohol EP)
prisma CHECK
(see "4.2
page 36).
Sterilization
®
Not permitted
EN
39