Detailed information about your system
9
System Accuracy Results for Glucose
Concentrations ≥5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within ±5%
Within ±10%
Within ±15%
72.7%
98.6%
100.0%
(314/432)
(426/432)
(432/432)
System Accuracy Results for Glucose
Concentrations between 1.2 mmol/L and
25.9 mmol/L
Within ±0.83 mmol/L or ±15%
99.5% (597/600)
NOTE: Where 1.2 mmol/L represents the lowest glucose
reference value and 25.9 mmol/L represents the highest
glucose reference value (YSI value).