Detailed information about your system
9

System Accuracy

Diabetes experts have suggested that glucose meters
should agree within 0.83 mmol/L of a laboratory
method when the glucose concentration is lower than
5.55 mmol/L, and within 15% of a laboratory method
when the glucose concentration is 5.55 mmol/L or
higher. Samples from 100 patients were tested using
both the OneTouch Select
 Plus System and the YSI 2300
®
Glucose Analyzer laboratory instrument.
System Accuracy Results for Glucose
Concentrations <5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within
Within
Within
±0.28 mmol/L
±0.56 mmol/L
±0.83 mmol/L
72.0%
94.6%
98.2%
(121/168)
(159/168)
(165/168)