5. Commissioning, operation
After approx. 2 seconds P1 appears in the first display line for the measured value 1,
dP appears in the 2nd line for the decimal point.
Use the arrow keys ▲▼ to select the desired number of decimal places:
EN
dP.
= two decimal places
dP . = one decimal place (decimal point moves one position to the right)
Press ENTER/MENU to confirm the desired setting. Si_ appears on the display. Use
the arrow keys ▲▼ to select the sign:
Si_ = the number to be entered is in the negative range (below 0.00 °C)
Si ┘ = the number to be entered is in the positive range
Press ENTER/MENU to confirm the desired setting. Fd 0 appears on the display.
Select the range using the arrow keys ▲▼:
Fd 0 = below 1,000 °C
Fd 1 = above 1,000 °C