1. Align the table mount so that it is parallel with the
table edge. The distance between the table edge
and the front holes on the table mount must be
12 cm.
2. Fasten the table mount to the table using the screws
supplied.
3. Lift the test unit slightly at the back.
4. Align the front feet on the test unit with the slots in
the table mount and move the test unit backwards in
the table mount. The back feet must engage with the
table mount.
5. Check that the high pressure pipe can be bled
manually.
WARNING
Risk of death or serious injury!
Improperly connecting a compressed air pipe can result
in death or serious injury.
► Compressed air pipes must only be connected by
trained personnel.
CAUTION
Risk of fire!
If the test unit is inflated with oxygen, there is a risk of
fire. Use only permissible pressures.
► Only breathing air as per EN 12 021 from a fixed
compressed air system or a compressed air cylinder
may be used for the test unit.
► The maximum permitted high pressure is 300 bar
(4500 psi). The maximum permitted medium
pressure is 20 bar (58 psi).
6. If necessary, connect the compressed air cylinder or
compressed air pipe to the high-pressure inlet using
an adapter for the high-pressure inlet (see "Order
list", page 20).
Instructions for use
|
Quaestor 6000/8000
12 cm
7. Connect the test unit to the PC with the USB cable.
NOTICE
The power plug must be freely accessible to ensure it
can be easily disconnected.
8. Connect the test unit to the power supply using the
power supply unit.
To disconnect the plug, pull the socket approx.
1 mm away from the test unit.
This releases the lock, allowing the plug to be
removed from the test unit.
5

Use

5.1 Preparations for use

5.1.1 Preparing the test unit
● Open the compressed air supply at the high-
pressure inlet before testing.
5.1.2 Start the test software on the PC
When a test is launched using the test software, the
test unit is switched on automatically. There is no on/off
switch on the test unit.
NOTICE
Do not connect any breathing safety product to the test
unit's low pressure measuring points before launching
the software.
Each time the software is started, a zero adjustment is
automatically performed before the first test.
CAUTION
Risk of crushing!
There is a risk of crushing to fingers when reaching into
the lung during operation.
► Only use a test unit with the test head attached.
1. Access the test software via the Start menu or the
desktop shortcut.
The startup screen appears. The program
components are loaded. If a system error is
detected, this is indicated by an error message.
A loading bar with status information provides
information on the progress of the startup process.
2. When the login screen appears, enter the user
name and password (see "Prior to first use",
page 15).
✓ After starting the program and successfully logging
on, the "Test due list" and "Test" tabs are displayed
in the function area.
|
Use
17