Commissioning the test unit
4.3.2 Fitting the brackets for the pressure
gauge or Bodyguard (Dräger
Quaestor 8000 only)
● Fit the bracket on the right or left under the test unit
with the two screws supplied.
4.3.3 Connecting the microphone (Dräger
Quaestor 8000 only)
1. Connect the supplied USB microphone to the PC.
2. Select and save the USB microphone in the test
software under Extras > Settings > Gauge
accuracy test.
3. Position the microphone near the warning device.
4.3.4 Fitting the adapter for mask/helmet
combinations
1. Use the drill supplied to drill a hole at each of the
points marked on the test head.
The top of the test head is approx. 5 mm thick; the
test head itself is hollow on the inside. The interior of
the test head cannot be damaged by the drilling.
16
2. Align the adapter and fasten it in place with the two
screws.
4.3.5 Setting up the test unit
Place the test unit on a table or workbench.
The PC should be placed next to the test unit so that it
is within reach of the operator.
NOTICE
Use only the original Dräger USB cable. A USB
extension cable is not permitted.
NOTICE
Use only the original Dräger power supply unit and
cable supplied.
Instructions for use
|