Indumax CLS50D/CLS50
10.4
Prozess
10.4.1
Prozesstemperatur
CLS50D
Sensorwerk-
CLS50D-*1/2
stoff
Ohne Flansch
PEEK
–20 ... 125 °C
(-4 ... 260 °F)
PFA
–20 ... 110 °C
(-4 ... 230 °F)
CLS50
Sensorwerk-
CLS50-*1/2
stoff
Ohne Flansch
PEEK
–20 ... 180 °C
(-4 ... 360 °F)
PFA
–20 ... 125 °C
(-4 ... 260 °F)
10.4.2
Prozessdruck (absolut)
max. 21 bar (305 psi), je nach Sensorausführung, s. Druck-Temperatur-Diagramm
Endress+Hauser
CLS50D-*3/4/5/6/8
DN50, ANSI 2"
–20 ... 125 °C
(-4 ... 260 °F)
–20 ... 110 °C
(-4 ... 230 °F)
CLS50-*3/4/5/6/8
DN50, ANSI 2"
–20 ... 180 °C
(-4 ... 360 °F)
–20 ... 125 °C
(-4 ... 260 °F)
CLS50D-*7
JIS
–20 ... 125 °C
(-4 ... 260 °F)
–20 ... 110 °C
(-4 ... 230 °F)
CLS50-*7
JIS
–20 ... 180 °C
(-4 ... 360 °F)
–20 ... 125 °C
(-4 ... 260 °F)
Technische Daten
CLS50D-*A/B/C
Losflansch PVDF
–20 ... 125 °C
(-4 ... 260 °F)
–20 ... 110 °C
(-4 ... 230 °F)
CLS50-*A/B/C
Losflansch PVDF
–20 ... 125 °C
(-4 ... 260 °F)
–20 ... 125 °C
(-4 ... 260 °F)
21