PR Series Balance
When pressing Tare/Zero
8.4

Service Information

If the troubleshooting section does not resolve your problem, contact an Authorized Ohaus Service Agent. Please
visit our website www.ohaus.com to locate the Ohaus office nearest you.
The weight on the pan is too heavy.
The pan is not properly installed.
The average piece weight of the
sample is too small (less than 0.1d)
in Parts Counting mode.
The reference weight on the pan is
too small (less than 100d) in
Percent Weighing mode.
System error
The environment is not stable.
The value is out of zero range.
The zero range is dependent on
region.
The tare value is a negative value.
System error
System error
Reduce sample size until the weight is
within the weighing capacity.
Install the weighing pan properly
Recaluate the APW value.
If error persists, please use a different
sample.
Put more samples on the pan
Please contact OHAUS service
(www.ohaus.com - More - Contact Us)
Move balance to a suitable location
Perform zeo again according to
regional requirement.
Press Zero instead of Tare
Please contact OHAUS service
(www.ohaus.com - More - Contact Us)
Disconnect the power and restart.
If error persists, please contact
OHAUS service (www.ohaus.com -
More - Contact Us)
EN-33