Cerabar S PMC71, PMP71, PMP75 mit FOUNDATION Fieldbus
Alarm
747
17
General Error
707
18
Calibration Error
711
18
Calibration Error
713
18
Calibration Error
721
18
Calibration Error
722
18
Calibration Error
723
18
Calibration Error
741
18
Calibration Error
719
19
Configuration Error
750
19
Configuration Error
122
20
Electronics Failure
101
20
Electronics Failure
716
20
Electronics Failure
725
20
Electronics Failure
704
20
Electronics Failure
703
20
Electronics Failure
705
20
Electronics Failure
737
20
Electronics Failure
738
20
Electronics Failure
739
20
Electronics Failure
742
20
Electronics Failure
744
20
Electronics Failure
743
20
Electronics Failure
748
20
Electronics Failure
113
20
Electronics Failure
Endress+Hauser
8.4
Diagnosemeldungen im DIAGNOSTIC Transducer
Block (TRDDIAG)
• Der Parameter Aktuelle Diagnose (actual diagnostics) zeigt die Meldung mit der höchsten Priori-
tät an. Jede Meldung wird zusätzlich gemäß FOUNDATION Fieldbus-Spezifikation über die Para-
meter XD_ERROR und BLOCK_ERROR angezeigt.
• Über den Parameter Diagnose kann man den aktiven Alarm mit der höchsten Priorität einsehen.
• Über den Parameter Letzte Diagnose kann man den letzten nicht mehr aktiven Alarm einsehen.
8.4.1
Erklärung XD_ERROR, BLOCK_ERROR und Verhalten der Aus-
gänge
0
Other
0
Other
0
Other
0
Other
0
Other
0
Other
0
Other
0
Other
0
Other
0
Other
7
Sensor Failure
0
Other
0
Other
0
Other
7
Sensor Failure
0
Other
0
Other
0
Other
0
Other
0
Other
0
Other
0
Other
7
Sensor Failure
7
Sensor Failure
0
Other
BAD_SENSOR_FAILURE
BAD_NON_SPECIFIC
BAD_NON_SPECIFIC
BAD_NON_SPECIFIC
BAD_NON_SPECIFIC
BAD_NON_SPECIFIC
BAD_NON_SPECIFIC
BAD_NON_SPECIFIC
BAD_NON_SPECIFIC
BAD_NON_SPECIFIC
BAD_SENSOR_FAILURE
BAD_SENSOR_FAILURE
BAD_SENSOR_FAILURE
BAD_SENSOR_FAILURE
BAD_DEVICE_FAILURE
BAD_DEVICE_FAILURE
BAD_DEVICE_FAILURE
BAD_DEVICE_FAILURE
BAD_DEVICE_FAILURE
BAD_DEVICE_FAILURE
BAD_DEVICE_FAILURE
BAD_DEVICE_FAILURE
BAD_DEVICE_FAILURE
BAD_DEVICE_FAILURE
BAD_DEVICE_FAILURE
Diagnose und Störungsbehebung
Pressure, Level, Flow
All
Level
Primary Value(1)
Level, Flow
Primary Value(1)
Totalizer 1 (6)
Level
Primary Value(1)
Level
Primary Value(1)
Level
Primary Value(1)
Flow
Primary Value(1)
Totalizer 1 (6)
Level
Primary Value(1)
Level
Primary Value(1)
Pressure, Level, Flow
Primary Value(1)
Pressure(3)
Maximum Pressure(4)
Counter P > Pmax(5)
Totalizer 1 (6)
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
Pressure, Level, Flow
All
77