4.
Troubleshooting
In instances of abnormal operation, error messages will be displayed on the Control
Panel. If a problem persists, please request assistance from the customer support
information services.
Error Messages
Error Code
E.02
E.03
E.04
E.05
E.06
E.07
E.08
EzRay Air W (Model: VEX-S300W) User Manual
Check Parameter
X-ray Generator
X-ray Generator
X-ray Generator
X-ray Generator
X-ray Generator
System
User
4. Troubleshooting
Description
The kV feedback is lower than the
set value during X-ray irradiation.
The kV feedback is higher than the
set value during X-ray irradiation.
The mA feedback is lower than the
set value during X-ray irradiation.
The mA feedback is higher than the
set value during X-ray irradiation.
The mono-tank block temperature
exceeds the error limit.
The Remote Exposure Switch was
pressed when the system power is
turned on.
1) The Remote Exposure Switch was
turned off earlier than the exposure
time.
2) The door of the X-ray room was
opened during X-ray exposure when
the door has the Door Interlock
Switch (option) installed.
35