Herunterladen Inhalt Inhalt Diese Seite drucken

Other Problems; Disposal - Omron HBP-1320 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für HBP-1320:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 82
HBP-1320-E main.book Page 29 Tuesday, May 8, 2018 5:16 PM

Other problems

Error
Priority
code
Inflated the cuff to 300 mmHg or more
during inflation in "Auscultation Mode".
E3
Low
Over-inflation occurs
E40
Medium The battery is depleted.
E41
Medium Battery failed to charge.
E42
Medium Battery voltage error

Disposal

Description
As there is a risk of environmental pollution, follow your applicable national and local legal regulations
regarding disposal or recycling of this equipment and batteries.
The main constituents of each part are listed in the table below. As there is a risk of infection, do not
recycle patient attachments such as cuffs, but dispose of them as instructed by your facility's
procedures and applicable regulations.
Item
Package
Main unit and
accessories
Battery pack
Cause
When inflating in "Auscultation Mode",
release the button when the pressure
reaches the desired value.
If this occurs during measurement, repeat
measurement.
If this occurs when not performing
measurement, contact your OMRON retail
dealer or distributor.
Recharge the battery, or replace with a new
battery. (page 14)
Try charging again. If continues to fail,
replace with a new battery. (page 14)
Replace the battery with a new battery. If the
error continues, contact your OMRON retail
dealer or distributor.
Parts
Box
Cardboard
Inner dividers
Cardboard
Bags
Polyethylene
ABS (Acrylonitrile butadiene
Enclosure
styrene), Polycarbonate,
Silicone rubber
Internal parts
General electronic components
Outer tube
Polyvinyl chloride
Cell batteries
Nickel-metal hydride
Internal parts
General electronic components
Solution
Material
EN
29

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis