Environmental Conditions

Operational temperature and
humidity
Storage and transportation
Non-Invasive Blood Pressure (NIBP)
Measurement technology
Measurement method
Pressure display range
Pressure display accuracy
NIBP measurement range
NIBP accuracy*
Pulse rate accuracy
Reference standard:
* Comparison with auscultation method performed by a trained professional.
DIA determined by the auscultation method is "K5".
Note:
Specifications may be changed without prior notice.
This blood pressure monitor fulfils the requirements of the EC directive 93/42/EEC (Medical Device
Directive). It also conforms to the European standard EN 1060, Non-invasive Sphygmomanometers
Part 1: General Requirements and Part 3: Additional Requirements for Electromechanical Blood
Pressure Measuring Systems.
Temperature range: 5 to 40°C (41 to 104°F)
Humidity range: 15 to 85%RH (not condensed)
Atmospheric pressure: 700 to 1060hPa
Temperature range: -20 to 60°C (-4 to 140°F)
Humidity range: 10 to 95%RH (not condensed)
Atmospheric pressure: 500 to 1060hPa
Oscillometric
Dynamic Linear Deflation method
0 to 300 mmHg
Within ±3 mmHg
SYS 60 to 250 mmHg
DIA 40 to 200 mmHg
Pulse rate 40 to 200 /min
Maximum mean error within ±5mmHg
Maximum standard deviation within 8mmHg
Within ±5 % of reading
EN1060-1:1995+A2:2009
EN1060-3:1997+A2:2009
EN80601-2-30:2010+A1:2015
EN ISO 81060-2:2013
EN
31