EN
Appendix A: Informations reg. EMF according to EN 60601-1-2
The b-intense MED are intended for use in the electromagnetic environment specified below. The customer or the user of the above listed
models should assure that they are used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
The b-intense MED are intended for use in the electromagnetic environment specified below. The customer or the user of the b-intense MED
should assure that they are used in such an environment.
Immunity test
Test Condition
Electrostatic
±,8 kV Contact
discharge (ESD)
± 2,4,8,15 kV Air
IEC 61000-4-2
Radiated RF EM
3 V/m
fields and Proximi-
80 MHz – 2,7 GHz
ty fields from RF
80 % AM 1 kHz
wireless communi-
cations equipment
IEC 61000-4-3
385 MHz (18Hz Pulse Modulation)
450 MHz (FM+/-5KHz deviation 1kHz
sine or 18Hz Pulse Modulation)
710 MHz (217Hz PM)
745 MHz (217Hz PM)
780 MHz (217Hz PM)
810 MHz (18Hz PM)
Compliance
Electromagnetic environment - guidance
Group 1
The b-intense MED use RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference
Class B
in nearby electronic equipment.
Enclosure Port
IEC 60601
Compliance level
±,8 kV Contact
± 2,4,8,15 kV Air
3 V/m
80 MHz – 2,7 GHz
27 V/m
28 V/m
9 V/m
9 V/m
9 V/m
28 V/m
60 60
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a profes-
sional healthcare facility environment.