Plugging the cabin in and out is only permitted by pulling the plug, which is connected to the cable.
Don't pull the cable itself, this could damage the cable and subsequently lead to an electric shock.
Using this device next to other devices should be avoided, for it could lead to incorrect operation. If
it is necessary, all affected devices should be monitored to secure regular operation.
To connect the electrical connection lines correctly, use the connection plan.
Use of this equipment adjacent to or stacked with other equipment should be avoided because it
could result in improper operation. If such use is necessary, this equipment and the other equipment
should be observed to verify that they are operating normally.
Use of accessories, transducers and cables other than those specified or provided by the manufactu-
rer of this equipment could result in increased electromagnetic emissions or decreased electromag-
netic immunity of this equipment and result in improper operation.
Portable RF communications equipment (including peripherals such as antenna cables and exter-
nal antennas) should be used no closer than 30 cm (12 inches) to any part of the b-intense MED,
including cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
If the Essential Performance is lost or degraded due to EM Disturbances, the b-intense MED will
switch off all infrared heaters and the b-control MED unit will switch to the Stand-by mode.
The treatment cannot be carried out.
The b-intense MED is only intended to be used in professional healthcare environment!
Visually check the b-intense
MED regularly. In case of visible damage unplug the b-intense
MED
®
®
from the power supply and contact us via office@b-intense.at.
37 37