b-intense
MED: transportation and storage
®
Transportation of b-intense
The b-intense
MED must be transported on pallets by a transport company suitable for pallet transport and fragile goods. In case of visible
®
damage to the packaging, please contact the b-intense
Storage of the b-intense
The storage of the b-intense
®
quality check quickly. The stored product must remain on the pallet until the final installation.
Transport and storage conditions
Ambient Temperature: 0-40 °C
Realtive humidity: 10-100 %
Atmospheric pressure: 50-110 kPa
Placement of the b-intense
The best location is close to a socket, which will also be easily accessible after the assembly of the cabin. Ensure that there are no tripping
hazards near the b-intense
MED.
®
Please consider the following recommendations for installation and storage:
› The cabin must be stored and installed indoors.
› The ambient temperature of the b-intense
› The optimal humidity should not exceed 60 %. If this is not possible, suitable ventilation of the space to be used must be arranged.
› The cabin must be protected from any direct water spray and from wet or damp from the floor. Condensation in the room where the cabin
is sited must be prevented and adequate ventilation must be ensured. Condensation on the components of the cabin must be avoided.
› Ensure that the floor surface for the cabin is flat and level. Minor irregularities in the surface can be offset by the adjustable feet fixed to
the underside of the cabin base.
› The space between the wall and the cabin of at least 2 cm has to be considered.
› The distance from the cabin roof to the ceiling above should be at least 10 cm.
The accessability of the mains plug must always be ensured for disconnecting from the mains network in an emergency.
Please observe the following safety-relevant parameters:
› The horizontal minimum distance of the back-heater to burning materials is 10 cm.
› The horizontal minimum distance of the front-heater to burning materials is 30 cm.
› The heaters must not be covered.
› The power circuit must be secured using a fault-current circuit breaker (FI) with a trip current of 30 mA.
Service and Maintenance
Annual preventive replacement (or replacement after 100 hours of operation) of two 10 A fuses inside of the main unit by
service personnel is recommended.
Service and maintenance has to be performed by service personnel. Patient and operator must not do any service and
maintenance. The expected life time of the system is 10 years.
MED
®
office immediately. The cabin must not be assembled until release.
®
MED
®
MED should be as short as possible and not exceed one month in order to be able to complete the b-intense
MED
®
®
MED should be at least 18 °C (room temperature) and at maximum 35 °C.
41 41