Operation
Start Up
With the ULS200 correctly installed (or aimed at a wall 0.25 to 5 m away), apply power and
view the start up sequence. It will light all possible LED values, show product revision number,
product model number, and will enter run mode. Then, the display shows the measurement of
the distance from the transducer face to the material level in the units indicated.
Display / Operation Status
output 1
output 2
operation status
Operation Status - Run Mode
- echoes are valid and within range.
- echoes are lost due to poor conditions or out of range. This may be typical in
applications where there are deep vessels and the material level is normally out of
range. Refer to Troubleshooting.
- extended loss of echo period. Operation has gone into fail-safe. Refer to
Troubleshooting.

Quick Start

To set a basic high/low application where you can easily adjust the measured distance, use
the method below. The Pointek ULS200 relays are preset as: relay 1 = alarm 1, high alarm at
0.25 m; relay 2 = alarm 2, low alarm at 5.00 m.
To change the setpoints by reference method, set the material or target to the distance as
displayed. Press the ` 1 /
value. Press the alarm key a second time so the ULS200 changes the setpoint to the value
currently being measured. After viewing or changing the setpoint, the Pointek ULS200 reverts
to the run mode.
relay 1
1.
Position the unit so that it reads 0.75 m.
2.
Press 1 /
3.
Press 1 /
4.
Once installed, the unit will register a high alaram at 0.75 m from the sensor face. If
Loss of Echo
Page EN-8
reading
units
'or ` 2 /
' key. The display shows the current setpoint function and
function display: high alarm
appears, complete steps 1 to 3 again.
Pointek ULS200 – QUICK START MANUAL
program
mode
↑ key
↓ key
current setpoint, 0.5 m
adjustment
value
units
7ML19981XB83