PERFORMANCE SPECIFICATIONS:
The table below shows Arms (Accuracy Root Mean Square) values measured using rainbow sensors with Masimo SET
Oximetry Technology in a clinical study.
SaO
versus error (SpO
2
2
All manuals and user guides at all-guides.com
70 - 100%
SpO
A
2
rms
90-100%
1.57 (1,57) %
80-90%
1.80 (1,80) %
70-80%
2.47 (2,47) %
70-100%
1.98 (1,98) %
– SaO
) with linear regression fit and upper 95% and lower 95% limits of agreement.
2
rainbow Series
60 - 80%
SpO
2
60-70%
70-80%
60-80%
rainbow Series
70 - 100%
SpO
+ SaO
2
2
2
60 - 80%
SpO
+ SaO
2
2
2
108
A
rms
3.42 (3,42) %
2.49 (2,49) %
2.99 (2,99) %
10571B-eIFU-0620