ENGLISH
STERILE (ETO) unless package has
been opened or damaged.
FOR SINGLE USE ONLY.
DESCRIPTION
VivaSight Endobronchial Blocker Tube
(EBB) is a single-use endobronchial
blocker designated for one-lung ventilation.
VivaSight EBB is used in combination
with Ambu® VivaSight™ 2 SLT or
equivalent Endotracheal tube. VivaSight
EBB features a blue silicone cuff.
VivaSight EBB can be deflated and
repositioned under visual control
provided by VivaSight 2 SLT. The angled
distal tip of VivaSight EBB can be
adjusted to facilitate navigation to the
bronchi.
Evacuation
port
Blocker tube
Bronchoscope
port
Anesthesia
circuit
INDICATIONS FOR USE
VivaSight EBB is used to isolate the left
or right lung of a patient in surgery, that
requires one lung ventilation or one lung
anesthesia.
CONTRAINDICATIONS
VivaSight EBB is contraindicated for use
in procedures involving the use of laser
equipment in the immediate area of the
device.
VivaSight EBB is not recommended for
use on patients with serious throat
edema or inflammation, hemorrhage, neck
vertebra trauma or wet lung.
Use of endobronchial blockers is
contraindicated in patients with aortic
aneurysm with tracheal compression.
WARNINGS
• VivaSight EBB is a single use product.
Valve
Do not reuse: After use, handle and
dispose according to local regulations
Pilot
for disposal of infected medical devices.
balloon
Reuse of this single use product may
create a potential risk to the user.
2-way
Reprocessing, cleaning, disinfection and
connector
sterilization may compromise product
characteristics which in turn create an
Twist lock
additional risk of physical harm to or
infection of the patient.
• Expert clinical judgment should be used
in selecting the suitable tracheal tube
Multi-joint
size for each patient. Use the same size
connector
VivaSight 2 SLT as the standard tube
size required in combination with
Tube
VivaSight EBB.
connector
• Make sure the cuff is completely
deflated prior to inserting or removing
VivaSight EBB.
• Do not attempt to resterilize any part of
VivaSight EBB.
• Prior to use, check that the pouch seal
Cuff
is intact and there are no impurities or
damage on the product such as rough
surfaces, sharp edges or protrusions
which may harm the patient.
- 3 -