CAUTION!
down for the installation or removal of the probe. If you use the quick connector,
the process pressure can be maintained during the installation or removal of the
sampling cell. Take a firm hold of the device to keep it in your hands while
removing it.
To connect to a pressurized process using DSC74:
1. Check that the DM70 pressure setting is correct (same as the process pressure). For
instructions, see
2. Select the quick connector or thread adapter that matches your process fitting.
3. Carefully seal the threads of the quick connector or thread adapter with PTFE thread seal
tape.
4. Attach the quick connector or thread adapter onto the sampling cell threads. Tighten the
fitting with a wrench.
5. Connect the sampling cell to the process fittings. Seal the fitting with PTFE thread seal
tape.
6. Install the gasket (delivered with the probe) to the nut of probe thread.
7. Set the probe into the sampling cell. Tighten the probe by turning it from the thread nut.
Do not tighten the probe from the handle.
16
If you use the thread adapters, the process pressure must be shut
Configuring pressure settings (page
13).
M212551EN-B