Aquatec® Ocean E-VIP

8 Troubleshooting

8.1 Identifying and Repairing Malfunctions

The following table provides information on
malfunctions and their possible causes.
If you cannot correct the malfunction using the
remedies named, please contact your provider directly.
Malfunction
An acoustic signal sounds and indicator
lamp D lights up while the device is
moving
The service indicator lamp F lights up
30
Possible Cause(s)
Battery almost exhausted
Hand control faulty
Software error
Emergency lowering mechanism button
faulty
Power failure while moving
Drive faulty
Controller faulty
D
C
B
A
G
H
Fig. 8-1
Remedy
Charge battery, refer to 4.12 Charging
Battery, page 20
Implement reset (see below table)
E
F
1556395-D

Werbung

loading