Respironics ComfortClassic Anleitung Seite 3

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

CIDEX
RESPIRONICS MASK
EXPOSURE TIME
Profile™ Lite
Simplicity™
ComfortClassic™
ComfortSelect™
Total™
CIDEX O.P.A.
RESPIRONICS MASK
EXPOSURE TIME
Simplicity™
ComfortClassic™
ComfortSelect™
Total™
RINSE
45 minutes
Per manufacturer's
instruction
45 minutes
Per manufacturer's
instruction
45 minutes
Per manufacturer's
instruction
45 minutes
Per manufacturer's
instruction
45 minutes
Per manufacturer's
instruction
RINSE
12 minutes
Per manufacturer's
instruction
12 minutes
Per manufacturer's
instruction
Per manufacturer's
12 minutes
instruction
Per manufacturer's
12 minutes
instruction
RESULT
NO. CYCLES TESTED
High level disinfection
High level disinfection
High level disinfection
High level disinfection
High level disinfection
RESULT
NO. CYCLES TESTED
High level disinfection
High level disinfection
High level disinfection
High level disinfection
NOTES
FOR ALL MASKS:
• Inspect the mask after
processing. If any components
are damaged, replace the mask.
• Slight discoloration of the mask
cushion after processing is normal.
30
• If the swivel does not move
freely after disinfection,
replace the split washer
(part #1002720 – pkg. of 5).
30
30
• If the swivel does not move
freely after disinfection,
replace the split washer
(part #1002720 – pkg. of 5).
30
• If the swivel does not move
freely after disinfection,
replace the sealing ring
(part #452027 – pkg. of 5).
10
• If the swivel does not move
freely after disinfection,
replace the sealing ring
(part #452027 – pkg. of 5).
• Verify function of the flapper
valve after processing. If the
flapper does not move freely,
replace the mask.
NOTES
FOR ALL MASKS:
• Inspect the mask after
processing. If any components
are damaged, replace the mask.
• Slight discoloration of the mask
cushion after processing is normal.
30
30
• If the swivel does not move
freely after disinfection,
replace the split washer
(part #1002720 – pkg. of 5). The
gel spacer is incompatible with
Cidex O.P.A. DO NOT DISIN-
FECT/STERILIZE THE GEL
SPACER. IT SHOULD BE
CLEANED ONLY.
(part #1002464).
30
• If the swivel does not move
freely after disinfection,
replace the split washer
(part #1002720 – pkg. of 5).
10
• If the swivel does not move
freely after disinfection,
replace the sealing ring
(part #452027 – pkg. of 5).
• Verify function of the flapper
valve after processing. If the
flapper does not move freely,
replace the mask.