Herunterladen Inhalt Inhalt Diese Seite drucken

Xylem SI Analytics TitroLine 7800 Gebrauchsanleitung Seite 201

Titrator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

The following calculation formulae are available for EQ and EP:
Formula for linear and dynamic
titration to EQ1
No formula
(EQ1-B)*T*M*F1/(W*F2)
(B–EQ1)*T*M*F1/(W*F2)
(B*F3–EQ1*F1)*T*M/(W*F2)
(W*F2)/(EQ1-B)*M*F1)
(W*F2)/(EQ1-B)*M*T*F1)
(W*F2)/(B-EQ1)*M*T*F1)
EQ1
The abbreviations used here have the following meaning:
ml:
Total consumption, e.g. for pH Stat
S:
Slope in ml/time (pH Stat)
EQ:
Consumption at the equivalence point 1 and 2 in ml
EP:
Consumption at the end point in ml
B:
Blank value in ml. Mostly determined by way of titration
T:
Titer of the titration solution (e.g. 0.09986)
M:
Mol; mol- or equivalence weight of the sample (e.g. NaCl 58.44)
F1 - F5
Factor 1 - 5 conversion factors
"Weight", weighed-in quantity in g or volume in ml
W
Formula for titrations to
End-Point (EP 1 and EP2)
(EP1-B)*T*M*F1/(W*F2)
(B-EP1)*T*M*F1/(W*F2)
(B*F3–EP1*F1)*T*M/(W*F2)
(W*F2)/(EP1-B)*M*F1)
(W*F2)/(EP1-B)*M*T*F1)
(W*F2)/(B-EP1)*M*T*F1)
EP1
EP2*T*M*F1/(W*F2)
(EP2-EP1)*T*M*F1/(W*F2)
(F3*EP2-EP1)*T*M*F1/(W*F2)
(F1/W) * EP1 *F2
((F1/W)*(EP2-EP1) * F3-F4)*F5
Information
No result will be determined.
Formula for calculating the
concentration of a sample taking into
account a blank value in terms of ml.
Direct titration to one EQ or EP1 (ex.:
chloride, p or m value)
Formula for calculating the
concentration of a sample taking into
account a blank value in terms of ml.
Reverse titration (examples. CSB,
saponification number)
Formula for calculating the
concentration of a sample taking into
account a blank value, including a
multiplicative factor.
Back titration.
Formula for calculating a titer (T)
of a titration solution.
Formula for calculating the
concentration of a sample taking into
account a blank value in ml.
Direct titration to one EQ or EP1.
Formula for calculating the
concentration of a sample taking into
account a blank value in ml. Back
titration (NCO-value, Epoxy-number).
Calculation of the consumption in the
equivalence or end point.
Formula for the calculation of
concentration of a sample. Direct
titration to 2 EP. Here EP2 (p and m
value)
Formula for the calculation of the
concentration of a sample. Direct
titration to 2 EP. Here calculation of the
difference between EP2-EP1.
Formula for the calculation of the
concentration of a sample. Direct
titration to 2 EP. Here: calculation of the
difference between EP2-EP1, taking
into account a multiplicative factor for
EP2.
Calculation of the des TAC (Total
Anorganic Carbonat reserve)
Calculation of the FOS (Volatile Organic
Acids)
FOS/TAC-value
201

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis