Herunterladen Inhalt Inhalt Diese Seite drucken

Installation; Incoming Acceptance, Transport, Storage; Transportation And Storage; Installation Conditions - Endress+Hauser RIA46 Kurzanleitung

Vorschau ausblenden Andere Handbücher für RIA46:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Installation

3
Installation
3.1

Incoming acceptance, transport, storage

The permitted ambient and storage conditions must be observed. The precise specifications can
be found in Section 10 "Technical data".
3.1.1 Incoming acceptance
On receipt of the goods, check the following points:
• Are the packaging or contents damaged?
• Is anything missing from the delivery? Compare the scope of delivery with the information
you specified in the order. See also section 2.2 "Scope of delivery".

3.1.2 Transportation and storage

Note the following points:
• Pack the device so that is protected against impact for storage and transportation. The original
packaging provides optimum protection.
• The permitted storage temperature range is -40 to +85°C (-40 to +185 °F); it is possible to
store the device in the limit temperature ranges for a limited period (maximum 48 hours).
3.2

Installation conditions

The device is designed to be used in the field
The orientation is determined by the readability of the display. Cable entries are located on the
bottom of the device.
Operational temperature range:
-40 to 50 °C (-40 to 122 °F)
#
Warning!
When using the field meter in explosion hazardous area observe the installation conditions given
in the corresponding safety notes.
"
Caution!
If the device is operated in the upper temperature limit range, this reduces the operating life of
the device.
!
Note!
Readability of the display cannot be guaranteed at temperatures below -30 °C (-22 °F).
1)
According to UL approval panel or surface mounting only.
28
1)
.
RIA46
Endress+Hauser

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis