Herunterladen Inhalt Inhalt Diese Seite drucken

Calibration Sub-Menu (Incal Models); Internal Calibration; Automatic Calibration - OHAUS Adventurer Analytical AX224N Handbuch

Vorschau ausblenden Andere Handbücher für Adventurer Analytical AX224N:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

ADVENTURER BALANCES
5.2 Calibration
Adventurer Balances (InCal models) offer a choice of six calibration methods: Internal Calibration,
Automatic Calibration, AutoCal™ Adjustment, Span Calibration, Linearity Calibration and Cal Test
(Calibration Test).
Note: The calibration unit is always in grams.
Attention: Do not disturb the balance during any calibration.
5.2.1

Calibration sub-menu (InCal models)

Internal
Automatic
Calibration
Calibration
Note: /E models only have Span Calibration, Linearity Calibration and Cal Test.
5.2.2

Internal Calibration

Calibration is accomplished with the internal calibration mass. Internal Calibration can be performed at
any time, provided the balance has warmed up to operating temperature and is level.
With the Balance turned ON and no load on the pan, touch Internal Calibration. Alternatively, press the
Cal key on the balance.
The Balance begins to calibrate. And no load on the pan, touch Internal Calibration.
The display shows the status, then returns to the current application.
To cancel at any time, press Save.
5.2.3

Automatic Calibration

When Automatic Calibration is set ON, the balance performs a self-calibration:
when it senses a temperature change of 1.5° C
or every 11 hours
AutoCal will automatically calibrate the Balance (using the internal mass) each time there is a change in
temperature significant enough to affect accuracy or every 11 hours.
An information screen will appear when an Automatic Calibration is about to start. Three option buttons
will be displayed:
– Press to perform the calibration directly.
Now
– Press to perform the calibration after 5 minutes.
5 min
Deactivate – Press to deactivate the Automatic Calibration function.
AutoCal™
Adjustment
Calibration
Span
Linearity
Calibration
EN-29
Cal
Test

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Diese Anleitung auch für:

Adventurer analytical ax124Adventurer analytical ax124/eAdventurer analytical ax224Adventurer analytical ax224/eAdventurer analytical ax324Adventurer analytical ax223 ... Alle anzeigen

Inhaltsverzeichnis