4.7 Display Hold

Note: Before using any application, be sure the balance has been leveled and calibrated.
Two Modes are available:
Display Hold - allows the user to capture and store a stable weight.
Peak Hold - allows the user to capture and store the highest stable weight.
Display Hold
4.7.1
Display Hold
1. In the upper left portion of the home screen, select Display Hold
2. Place the sample on the pan and press Hold at any time while the weight is being captured.
3. The Main Display Line now shows the first stable weight.
4. Press Clear to remove the hold and return to Display Hold Home screen.

Peak Hold

4.7.2
Peak Hold
1. In the upper left portion of the home screen, select Display Hold
2. Choose Peak Hold Mode in Item Settings (see section 4.7.3).
3. Place sample on the pan and press Start to begin.
4. Continue to weigh samples. The highest stable weight will be held.
5. To remove the hold and return to normal operation press Stop.
The Display Hold Home screen
Main Display Line
Reference Fields
Function Buttons
EN-23