ADVENTURER BALANCES

USB INPUT

The balance will respond to various commands sent via the interface adapter.
Terminate the following commands when with a [CR] or [CRLF].
Adventurer Commands
zC
perform span calibration
0S
print unstable data
1S
print stable data only
P
same as pressing Print
SP
print stable weight only
IP
immediate print of displayed weight (stable or unstable)
CP
Continuous print of weights
SLP
Auto-print stable non-zero weight only
SLZP
Auto-print stable non-zero weight and zero reading
xP
Auto-print on 1 to 3600 second intervals (x = 1 to 3600)
0P
Ends interval print
T
same as pressing Tare
Z
same as pressing Zero
PV
print software version
Auto-Print Operation
Once Auto-Print is activated in the menu, the balance will send data as required.
If there is data in the print buffer the printer will finish printing this data.
10.

SOFTWARE UPDATES

Ohaus is continuously improving its balance software. To obtain the latest release, please contact your Authorized