Complications regarding continuous analgesia
Vascular injury, nerve damages, paresthesia, pain, failed block, motor deficits,
spread of local anaesthetic into the epidural area, infection
If complications occur while using the device, follow the protocols of your
organisation. If this does not resolve the complications, or if they are
regarded as serious or untreatable, carefully stop the procedure and remove
invasive device components from the patient.
Warnings
for sterile product:
This is a disposable medical device for use with only one patient.
This device must not be re-used under any circumstances.
This device must not be resterilised under any circumstances.
The materials used in the manufacture of this device are not suitable for repro-
cessing or resterilisation.
This device is not designed to be reprocessed or resterilised.
Unauthorised re-use or reprocessing
• can cause the device to lose the performance properties intended by
the manufacturer.
• leads to a significant risk of cross-infection/ contamination as a result of
potentially inadequate processing methods.
• may cause the device to lose significant functional properties.
• may cause materials to break down and lead to endotoxic reactions
caused by the residues.
regarding puncture
1. Take care to use devices of suitable dimensions (diameter, length), espe-
cially when treating obese patients and children.
2. For cannulas with stylets: Only perform the puncture (even when remov-
ing the cannula) with the stylet in place.
3. To avoid bending the cannula tip, never apply excessive force to the can-
nula
4. If you unexpectedly come into contact with bone, change the direction of
the cannula. Do not try to overcome bone resistance. Failure to adhere to
these rules could cause the tip of the cannula to bend.
5. Repeated bone contact will damage the cannula tip. On no account should
you continue to use a cannula damaged in this manner.
for catheter placement and removal:
1. Check that the catheter will pass through the cannula immediately before use.
2. The tip of the cannula can be damaged by bone contact during insertion.
12