5. Commissioning, operation / 6. Configuration
Medium temperature outside of the span
If the medium temperature exceeds that configured within the
transmitter an error will be signalled.
EN

6. Configuration

Configuration is carried out using a USB interface with a PC via
the model PU-548 programming unit (accessories, order number:
14231581). The connection with the thermometer is made via the
appropriate adapter cable (accessories: M12 x 1 circular connector,
order number: 14003193).
Measuring range, dampening, error signalling, Tag no. and other
parameters can be adjusted (see configuration software).
The measuring range is configurable between -50 ... +250 °C
[-58 ... +482 °F]. The configuration software checks the required
measuring range and will only accept permissible values. Intermediate
values are configurable - the smallest increment is 0.1 °C or 0.1 °F. The
thermometers are delivered configured to customer specifications within
the configurable limits.
Please note:
The measuring range of the thermometer is limited by the application
range of the measuring element, not by the setting range of the
transmitter.
18
Easy to use
LED status display
Compact design
No further voltage supply is needed for either
the programming unit or for the transmitter
(replaces programming unit model PU-448)