Herunterladen Inhalt Inhalt Diese Seite drucken

Service - Interco Skippy Gebrauchsanweisung

Therapiestuhl
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

Checking / components
Visual inspection of the seating system and the undercarriage, the attached
components and accessories for visible damage
Check of connection between undercarriage and seating system
Check working condition of the brakes
Visual inspection of all fastening elements
Visual inspection of all accessories / options
Retightening of all fastening elements
Retightening of all screw connections of the accessories / options
Performance test seat tilt system
Performance test seat height adjustment
11.3.

Service

Test / Execution
Cleanliness
Check of all settings of the SKIPPY
Check of setting of the head rest:
Does the head rest lie against the patient's occiput (standard version)?
Does the bottom edge of the head rest lie against the base of the head (shell shape)?
Do the side elements exert pressure on the patient's head?
Does the patient have sufficient freedom of movement?
Check of setting of the foot rest:
Is there enough free space at the back of the knee?
Check of setting of the arm rests:
Are the shoulders in a neutral / ergonomic position when the elbows are
in the resting position?
Check of setting of the lateral side supports::
Is there enough free space under the armpits?
Is the upper body guided sufficiently?
Check the safety distances according to DIN EN 12182:2012 using the following table
Safety distances according to DIN EN 12182:2012 for children
To avoid
Trapping fingers
Trapping feet
Trapping the head
Trapping the genitals
Safety distances according to DIN EN 12182:2012 for adults
To avoid
Trapping fingers
Trapping feet
Trapping the head
Trapping the genitals
Note!
Expiry of the manufacturer's warranty if service is not conducted
The product must be checked regularly - at least once a year - as part of a (payable) service� Otherwise
interco GmbH cannot provide any warranty�
therapy chair
®
Between moving parts
<4mm or >25mm
<25mm or >120mm
<60mm or >300mm
<8mm or >75mm
Between moving parts
<8mm or >25mm
<35mm or >120mm
<120mm or >300mm
<8mm or >75mm
Daily
Weekly
x
x
x
x
x
x
x
x
x
OK
Adjustments
necessary
Between fixed parts
<5mm or >12mm
<25mm or >45mm
<60mm or >250mm
<8mm or >75mm
Between fixed parts
<8mm or >25mm
<35mm or >100mm
<120mm or >250mm
<8mm or >75mm
MAINTENANCE PL AN
49

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis