shown. Do not depress Tap-On™ Foot Pedal while
installing batteries.
The communication light will blink for approximately
two seconds to indicate the Tap-On™ Foot Pedal's
ability to communicate with the unit. If the light does
not blink, check the batteries. If the batteries are good
and the light doesn't blink, a communications error
may exist. To re-establish communication with Tap-
On™ Foot Pedal review Synchronization procedure,
section 8.10.
The remote frequency communication can be
bypassed using the auxillary Tap-On™ Foot Pedal
cable. Refer to Section 12.2 Technical Support and
Repair for further action.
Replace the battery cover and screw and hand
tighten cover with philips screwdriver.
Remove batteries if Tap-On™ Foot Pedal is to be
stored for an extended period of time.
8.10 Tap-On™ Foot Pedal Synchronization
The Tap-On™ Wireless Foot Pedal supplied with your
system has been factory synchronized with the base unit.
Should a replacement Tap-On™ Foot Pedal be necessary,
synchronization will be required prior to system operation.
Perform the following steps to synchronize the Tap-On™
Foot Pedal with the base unit.
1.
Turn the MainPower switch located at the center
front underside of the system to the OFF (O)
position.
2.
Install a new set of "AA" batteries into the foot
control (see section 8.9). Leave the battery cover
of the Tap-On™ Foot Pedal open so the red push
button is accessible.
3.
Maintain a distance of no more than 10 feet (3
meters) between the base unit and Tap-On™ Foot
Pedal during the synchronization process.
4.
Remove any inserts from the handpiece and adjust
the Power Level Control out of Rinse Mode. Turn
the Main Power switch to the ON (I) position and
wait for the Diagnostic Display graphics to light
(refer to Section 9.2).
5.
While all graphics are lit, press the Purge button,
located on the Diagnostic Display. The graphics
will begin to blink in a sequential pattern,
representing the synchronization mode. This mode
will last 5 to 6 seconds.
PURGE
Look for blinking
communications light.
6.
During this mode, press the red button located in
the battery compartment of the Tap-On™ Foot
Pedal. This will complete the synchronization
process.
7.
Synchronization is successful when all graphic
lights blink at the same time.
8.
To verify proper communication, press the foot
control to the Boost position (Tap-On™ Foot Pedal
fully depressed – 2
nd
Boost graphic on base unit illuminates.
9.
Attach battery cover and tighten the screw.
10.
In the event communication cannot be established,
temporarily use the supplied Auxillary Tap-On™
Foot Pedal Cable to connect the Tap-On™ Foot
Pedal directly to the unit.