Additional specifications:
The measured value is obtained and displayed as a true effective value (TRUE RMS). Selectable
coupling type: AC or AC+DC. For the coupling AC+DC an additional error of 1 % must be taken
into account. In the case of non-sinus-shaped curves the display value becomes less precise.
Thus an additional error results for the following crest factors:
Crest factor from 1.4 to 2.0 additional errors + 1.0 %
Crest factor from 2.0 to 2.5 additional errors + 2.5 %
Crest factor from 2.5 to 3.0 additional errors + 4.0 %
HFR high-frequency suppression (low-pass filter):
AC accuracy plus 1 % of the measured value, 40 Hz to 400 Hz
Limiting frequency (- 3 dB): 800 Hz
Attenuation: approx. - 24 dB
Peak-Hold: ± 3 % of the measured value + 200 digits, 40 Hz to 1 kHz, sinusoidal
7.3 Resistance measuring ranges
Overload protection: 1000 V
Measuring range
400.0 Ω
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ
Max. no-load voltage: approx. 2.5 V
Max. short-circuit current: approx. 0.1 mA
7.4 Diode testing
Overload protection: 1000 V
Measuring range
2.000 V
Max. no-load voltage: approx. 2.5 V
Max. short-circuit current: approx. 0.1 mA
7.5 Continuity testing
Overload protection: 1000 V
Measuring range
400.0 Ω
Max. no-load voltage: approx. 2.5 V
Max. short-circuit current: approx. 0.1 mA
The integrated buzzer sounds at a resistance R lower than 30 Ω (preset). The resistance value
can be set within a range from 10 Ω to 50 Ω.
7.6 Capacity ranges
Overload protection: 1000 V
Conditions:
Capacitors discharged and connected in accordance with the polarity stated.
Measuring range
40.00 nF
400.0 nF
4.000 µF
40.00 µF
400.0 µF
4.000 mF
40.00 mF
7.7 Frequency ranges
Measuring range
400.0 Hz
4.000 kHz
40.00 kHz
100.0 kHz
12/ 2018
AC/DC
Resolution
0.1 Ω
1 Ω
10 Ω
100 Ω
1 kΩ
10 kΩ
AC/DC
Resolution
1 mV
AC/DC
Resolution
0.1 Ω
AC/DC
Resolution
Measuring time
0.01 nF
0.1 nF
1 nF
10 nF
100 nF
1 µF
10 µF
Resolution
0.1 Hz
1 Hz
10 Hz
100 Hz
BENNING MM 12
Measurement accuracy
± (0.2 % of measured value + 3 digits)
± (0.2 % of measured value + 2 digit)
± (1.0 % of measured value + 2 digit)
± (2.0 % of measured value + 25 digits)
Measurement accuracy
± (1.5 % of measured value + 3 digits)
Measurement accuracy
± (0.2 % of measured value + 3 digits)
1 s
± (1.0 % of measured value + 20 digits)
1 s
± (1.0 % of measured value + 10 digits)
1 s
1 s
± (1.0 % of measured value + 2 digits)
1 s
4 s
± (1.0 % of measured value + 10 digits)
8 s
± (1.0 % of measured value + 20 digits)
Measurement accuracy
± 3 digit for a 3¾-digit display
± 10 digit for a 4¾-digit display
Measurement accuracy