Measurement of SIN/COS signals
Adjustments and corrections (only MiniCODER plus)
If the measured results are not as expected or as required, you can make the follow-
ing adjustments with the aid of the buttons:
●
Ampl. auto adjust. -/+: Manual step-wise reduction/increase of the signal ampli-
tudes (in steps of approx. 50 mV)
●
Ampl. auto adjust. 0: Automatic optimisation of the amplitude synchronism relative
to the sine track (Vpp COS ≈ Vpp SIN)
●
Offset auto adjustment: Automatic minimisation of the offset on both signals
●
Phase auto adjustment: Automatic minimisation of the phase error between the two
signals (Δφ ≈ 0°)
GEL 211CS0
Sensor calibration group
7 Functions of the GEL 211C