CHANGING PRESSURE
• The default pressure of 50mmHg can be adjusted by pressing the SET button: The left display shows C1
(Chamber 1) and the right display shows a flashing air pressure of 50. Press the button once to advance the
pressure by increments of 10 from the default setting of 50mmHg up to a max of 120mmHg. Continuing to
press the button past 120 will reset the pressure to 00 and continue to advance in increments of 10. When you
reach the desired pressure, hold the SET button to select it. Repeat the steps to adjust the pressure in each
chamber.
NOTE:
• The light above the On/Off button will hold during inflation, and flash during deflation.
• When the device is off, holding the On/Off button 5 seconds restore the factory settings, meaning the total use
time will be erased and the pressure will be reset to the default 50mmHg for all four chambers.
• The device will remember the last set pressure and the total use time. When the device turns on, the total use
time shows on the left display.
• The maximum cumulative time is 99 hours. After 99 hours, it will reset to 0. If the total use time is less than 1
hour, it will display 00.
• C1, C2, C3, and C4 represent Chamber 1, 2, 3, and 4, respectively, while CH represents all 4 chambers together.
• The battery icon will flash during charging and will hold when fully charged.
• The device may be used while it is charging.
REMOVAL AND STORAGE
• Turn the device off before removing it from the leg.
• When not in use, do not store items in direct sunlight.
• When not in use, we suggest charging the device every three months.
CLEANING AND MAINTENANCE
NOTE: Inspect the device and follow the cleaning and disinfecting procedures prior to each use.
WARNING: Device must be turned off and disconnected from the wall outlet prior to and during cleaning or
disinfecting.
WARNING: Do not immerse device in any liquid for any reason. Do not place device in autoclave.
• Clean the outer surface of the pump unit using a soft cloth, moistened with soapy water or 70% isopropyl
alcohol. Air dry only.
• Clean the exterior of the cuffs using a soft cloth, moistened with soapy water or 70% isopropyl alcohol. Air
dry.
• Unit must be completely dry prior to use. To ensure that, leave the device in the OFF position and
disconnected from the wall outlet for at least 30 minutes (and as long as necessary for the unit to dry
completely) after cleaning or disinfecting.
• Do not remove the pump unit from the cuff.
• Do not place cuffs in dryer or microwave.
• Do not use hair dryer to accelerate drying.
• Do not place the device on top of or in front of portable or stationary radiators to accelerate drying.
• Do not use abrasive cleaners.