STEP-BY-STEP INSTRUCTIONS
6.3.5

Store your custom settings

To store your new settings for later recall, press either of the two Preset
keys and hold for approximately two seconds. The specific Preset key's
LED will illuminate when the settings have been successfully stored.
Settings for both the Pack side and Flow side are now stored in memory.
6.3.6

Preset Use

The two Preset buttons store all of the settings for both the Pack and Flow
sides of the device. The factory default settings are: 200°C temperature
and 40% tone volume for the Pack side and 170°C temperature and 60%
flow rate for the Flow side. The Thermal Response key is set at 90°C and
is fixed at that setting for the Pack side. The presets can be manually
programmed to store your personal settings by pressing and holding
either Preset button for at least two seconds. The Preset button's LED will
illuminate indicating that it is active.
6.3.7

Operation - PACK

Press the circumferential cuff on the hand piece to begin heating its tip to
the selected temperature setting. If the Temperature LED is illuminated,
the selected temperature will display on the numerical LED window. As
the tip's temperature rises, one tone frequency (approximately 2500Hz) is
heard. Once the tip is within 20°C of the selected temperature setting, the
tone will change to a lower frequency (approximately 2000Hz) and remain
at this tone until the hand piece cuff-switch is released. When pressing
the cuff, the tip will heat for a maximum time of 15 seconds if the selected
temperature setting is 200°C or lower. For temperature settings above
200°C, the tip will heat for no more than 10 seconds before timing out and
turning off. When in the thermal response mode, the tip will heat for up to
one minute while the cuff switch is depressed. To restart heating after a
timeout has been reached, release the hand piece cuff and then press it
again.