Ultra-Pro II™ Needle Guidance System
11. Select appropriate size needle guide insert and slide into position.
QUICK RELEASE FUNCTION
1. Press tab toward bracket to activate needle guide quick-release function.
REPROCESSING
WARNING
Users of this product have an obligation and responsibility to provide the highest degree of infection control to
n
patients, co-workers and themselves. To avoid cross-contamination, follow infection control policies
established by your facility.
Dispose of single-use components as infectious waste. Clean and disinfect reusable components after each
n
use.
See your ultrasound system's user guide for reprocessing transducer between use.
n
This procedure for reprocessing has been validated for effectiveness and compatibility. Device could be
n
damaged or cross-contaminated due to improper reprocessing.
Do not gas sterilize or autoclave bracket.
n
If bracket damage is evident, discontinue use and contact your CIVCO representative.
n
DISASSEMBLING BRACKET FOR CLEANING
NOTE: For multi-angle brackets only.
CLEANING THE BRACKET
1. After each use, remove bracket from transducer, discarding single-use, disposable components. Remove visible
contaminants from bracket surface using a small, soft instrument brush. Keep bracket from drying out until
complete cleaning can be accomplished.
2. Inspect bracket for damage such as cracks or breakage.
8.5FR,14,15,16,17,18,20,21,22,23GA
2.9,2.1,1.8,1.6,1.5,1.2,.9,.8,.7,.6mm
7
English