GCE MediVital E Bedienungsanleitung Seite 8

Druckregler mit flaschenventilen
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

2D code
EN
Empty display
Residual pressure indication
Security code
A.
The 2D code is intended for connecting the E-GAUGE
with a mobile application more easily.. Using Bluetooth
communication. The 2D code keeps information about
MAC address of E-GAUGE for filling stations or maintenance
purposes.
B:
Indication of "cylinder volume" parameter related to the
E-GAUGE setup. If the symbol ??L is displayed, it is
necessary to perform parameterization using the mobile
application. See par. 6.1.3
Conditions applying to the 2D code appearance: Measured
pressure < 2,5 bar (respectively 7 bar in the case of a new
valve)
Exit from the 2D code mode: As soon as the pressure
grows, the E-GAUGE switches to the standard display
(pressure measuring, remaining time calculating)
If the ambient temperature drops below -35°C, the display
operation is terminated automatically. This function
protects against any incorrect indication of characters due
to the low temperature. The display is switched on again
when the temperature rises above -30°C. This function is
not related to the condition of the battery.
Signalization of an incorrect RPV function
A.
Symbol „R" appears with a 30 min delay after the pressure
in cylinder drops below 2 bar.
Symbol „R" disappears if actual pressure in the cylinder is
A
above 10 bar.
If the parameters "cylinder volume" and "pressure unit" are
modified using the mobile application, the new values can
be saved by inserting the 4-digit security code generated
by the E-GAUGE display into the input field in the mobile
application (see. 6.1.3).
8/108