5.1.1.2. General description
The E-GAUGE is used to show information about the pressure inside the gas cylinder and about
the remaining time of the gas supply - until the complete exhaustion of gas cylinder - at the
actually set flow (flow head outlet use only).
B
A
5.1.1.3. Storage and operational conditions
Storage and operational conditions are in compliance with the conditions for the complete
MediVital®E combination valve mentioned in the chapter 3.
5.1.1.4. Display description
Indications during operation of the flow control head.
Indications when null flow is set or when the quick coupler is used
A – ELECTRONICS
The electronics of the E-GAUGE is located in the
plastic guard. The LCD display is used to indicate
of the operational information.
B – SENSOR OF THE FLOW CONTROL HEAD
POSITION
C
Inside the control unit there is an integrated
sensor of the flow control head position. It is
intended to detect the actual set flow to calculate
the time of the gas supply.
C – PRESSURE SENSOR
The pressure sensor installed in the valve port
provides a signal about the actual cylinder
pressure.
A.
Remaining time until the cylinder pressure drops to 15
A
bar at an actual flow. The indicated parameter is regularly
updated. The remaining time is automatically compensated
by the variation of ambient temperature.
B
B.
Bar chart equal to the pressure inside the gas cylinder.
C.
Actual pressure in the gas cylinder.
A
7/108
EN