5.1.2. VISUAL AND ACOUSTIC WARNINGS
Warning alarms are divided into two groups:
5.1.2.1. Operational alarms - related to the general use of the device. These are marked with a
warning triangle in the top part of the display.
Low pressure - first warning
EN
Low pressure - second warning
If the "Low pressure" alarm is shown, the remaining gas in the cylinder can be consumed
very quickly. Pay close attention to the cylinder, prepare a reserve gas supply, and possibly
discontinue the cylinder from service immediately.
Indication of the simultaneous use of a quick coupler and hose connection
In case the cylinder pressure drops under 50 bar.
The display background becomes black.
The warning symbol is indicated
The acoustic signal starts simultaneously with the flashing
red LED diodes located on the left of the display.
In case the cylinder pressure drops under 15 bar.
Information about the remaining time is exchanged by the
actual cylinder gas pressure value.
The acoustic warning signal starts simultaneously with
the flashing the red LED diodes located on the left of the
display.
When
the
E-GAUGE
consumption (about 4 l/min higher) than corresponding to
the flow outlet consumption only.
Only the pressure parameter is displayed
The warning symbol of QC (quick coupler) is indicated
If the QC outlet flow is too low, this status may not be
detected.
After the cylinder had been re-filled, the use of QC
quick coupling may be temporarily indicated during the
subsequent cooling down process.