Tempering of the device
5.1.2.2. Error alarms - related to the evaluation of the error status of the device. These are
marked with a big warning triangle in the left part of the display (except the "Low battery
voltage" warning)
Low battery charge
If MediVital®E is not used for long periods, check the display of the E-GAUGE every 3 months.
If the symbol
is shown on the display, the valve can still be used, but the valve must
be replaced at the latest within 3 months of detecting this situation. If this is not done, the
operation of the E-GAUGE may be unexpectedly terminated.
Error of the pressure measure
The warning is indicated if the E-GAUGE indicates a
rapid change in temperature caused by eg. moving the
device between two rooms with a high difference in
temperatures.
Only the pressure parameter is indicated.
The warning T symbol is indicated (temperature).
A warning is indicated when there is a low battery charge
(2,7V and less).
The battery symbol is permanently indicated
The battery symbol is static, it does not have an active bar
chart indicator and always includes one bar.
The use of the device may continue but at the earliest
cylinder filling, the battery must be changed.
The symbol may be indicated even when the battery
charge is OK, but the cylinder was moved to a room with
low ambient temperature. The symbol disappears when
the cylinder warms up again.
A warning is indicated when there is a defect of pressure
sensor.
The symbol ??? is indicated instead of the pressure or
remaining time.
The PS (pressure sensor) is indicate.
The bar chart is not indicated.
Discontinue the cylinder from service immediately and
send it back to the owner of the filling station. Service
must be performed by a GCE certified service centre.