Troubleshooting
Error
RFID -
error
524340
RFID -
error
524341
Remote
pump - test
failed
Remote
pump - lost
Remote
pump -
blocked
Remote
pump -
service
required
Time -
reset
Data log-
ger - error
65551
Data log-
ger - error
65548
Micro-
Tubes -
error
524351
Error -
error XYZ
Error
524320 -
self-test
failed
Error
524329 -
self-test
failed
54
Cause
Password for the RFID tag on
MicroTubes incorrect
Password for the RFID tag not avail-
able on MicroTubes
Leak test for the X-am pump and
accessories (e.g. hose, sensor) was
not successful
Communication with the X-am
pump interrupted during measure-
ment
X-am pump or accessories (e.g.
hose, sensor) blocked during mea-
surement
The X-am pump is in error mode
The internal time of the X-act 7000
has been reset
Error occurred when storing the
measurement result
Error occurred when reading the
data logger
Chemosensor in the MicroTubes is
already heavily discolored before
measurement
Unknown error occurred
The real time clock could not be cor-
rectly accessed during the self-test
The integration mechanism could
not be correctly moved during the
self-test
Remedy
MicroTubes can no longer be
used; insert new MicroTubes for
further measurements
MicroTubes can no longer be
used; insert new MicroTubes for
further measurements
Make sure that the X-am pump
and the accessories (e.g. hose,
sensor) are sealed within
45 seconds upon instruction
during the leak test; check the
hose for brittleness/leaks
Make sure that the coupler and
X-am pump are fitted correctly
on the X-act 7000
Make sure that the X-am pump
and accessories (e.g. hose, sen-
sor) are not blocked; when using
pre-tubes (e.g. ppb Booster
Basic), make sure that all tips
have been broken
Contact Service
Reset the time; measurements
are still possible
Turn the X-act 7000 off and back
on; check whether the error
occurs again during the next
measurement
Turn the X-act 7000 off and back
on; check whether the error
occurs again
MicroTubes can no longer be
used; use new MicroTubes for
further measurements
Restart the X-act 7000 and
repeat the process
Repeat the self-test, check the
displayed time and adapt as
necessary
Repeat the self-test
|