Dräger X-act 7000 Technisches Handbuch Seite 53

Vorschau ausblenden Andere Handbücher für X-act 7000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Error
Pneumat-
ics - error
524333
Pneumat-
ics - error
524334
Pneumat-
ics - error
524335
Pneumat-
ics - error
524339
Mechanics
- error
524330
Mechanics
- error
524331
Mechanics
- error
524350
Mechanics
- error
1048577
Mechanics
- error
1048578
Optics -
error
2097175
|
Technical Manual
Cause
Self-test of the pneumatic system
when starting the X-act 7000 or
before a measurement was not suc-
cessful; the pump was not able to
achieve the normal mass flow on
time due to a blockage in the gas
path;
The check to ensure that the Micro-
Tube to be measured is intact
before measurement was not suc-
cessful; the tip of the MicroTube
was damaged before measurement
The MicroTubes-specific mass flow
could not be achieved on time at the
start of the measurement due to a
blockage in the gas path; the tip of
the MicroTube on the outlet side
was not broken
Deviation of the mass flow drawn in
by the pump too large over the
entire measurement duration
Mechanical integration of the Micro-
Tubes into the gas path was not
successful; the outlet side of the
MicroTube could not be contacted
by the gas needle
Mechanical integration of the Micro-
Tubes into the gas path was not
successful; the inlet side of the
MicroTube could not be contacted
by the gas needle
MicroTubes could not be removed
After measurement, the MicroTubes
could not be disconnected from the
gas path; transport to the next,
unused MicroTube is not possible
After measurement, the MicroTubes
could not be disconnected from the
gas path and/or the MicroTubes
cannot be transported
Competing optical parameters
occurred (lighting duration and
image capture frequency); incorrect
optical parameters read out on the
RFID tag of the MicroTubes
®
X-act
7000
Troubleshooting
Remedy
make sure that the gas outlet is
not sealed (e.g. by fingers, con-
tamination)
Start new measurement
Make sure that the gas inlet or
outlet is not sealed (e.g. by fin-
gers, contamination); when
using a pre-tube (e.g. ppb
Booster Basic), make sure that
all tips have been broken
Make sure that the gas inlet or
outlet is not sealed (e.g. by fin-
gers, contamination); when
using a pre-tube (e.g. ppb
Booster Basic), make sure that
all tips have been broken
Start a new measurement; check
the MicroTubes housing for
excessive deformation and care-
fully bend it in the opposite direc-
tion if necessary
Start a new measurement; check
the MicroTubes housing for
excessive deformation and care-
fully bend it in the opposite direc-
tion if necessary
Try to remove the MicroTubes
once again; if necessary, care-
fully open the door in front
Turn the X-act 7000 off and back
on
Make sure that the wells/doors
for MicroTubes are not blocked;
re-start the device and carry out
the process once again
Do not use the MicroTubes any
further; insert new MicroTubes
for further measurements
|
en
53