MAINTENANCE

MAINTENANCE SCHEDULE

Notice!
The operating entity is obliged to ensure that all tests of the equipment are carried
out repeatedly at least once within every 24 months (EN 62353) or in intervals as
specified by the applicable national legal regulations. A report must be prepared on
the results of the tests (e.g.: according to EN 62353, Annex G), including the
measurement methods used.

Service interval signalization

Safety valve check

Downloaded from
www.Manualslib.com
- Unscrew the 4 screws from the rear of the main cover and 2 screws from
the rear part of the rail
- Disconnect the grounding wire
- Remove the main cover
- Reassemble using the opposite order
Maintenance
Clean air intake filter
Safety valve check
Replace filters in filter
Test the tightness of
joints and inspect the
equipment
Replace piston
complete with piston
rings, O-rings and
bearing
Replace suction filter
Perform "Repeated
Test" according to
EN 62353
Once 4000 operating hours have been reached, the value "4000" starts to
appear as the outlet pressure from the equipment, alternating with the actual
outlet pressure value. They alternate about every 2 seconds. Once this
information appears, the operator is obliged to secure service work for the
equipment.
The safety valve is adjusted to 7 bar by the manufacturer, then tested
and marked. It must not be readjusted.
manuals search engine
Page
Time interval
13
At least once a week
15
Once per year
16
Every 4000 hours
15
After two years
Service
Every 8000 hours
documentation
17
Every 8000 hours
14
1 x 2 years
16
To be
performed by
Staff
Qualified expert
Qualified expert
Qualified expert
Qualified expert
Qualified expert
Qualified expert