Werbung

Important EMC notices for use in the medical environments
• The UP-DF550 needs special precautions regarding
EMC and needs to be installed and put into service
according to the EMC information provided in this
instructions for use.
• The portable and mobile RF communications equipment
such as cellular phones can affect the UP-DF550.
The UP-DF550 is intended for use in the electromagnetic environment specified below.
The customer or the user of the UP-DF550 should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Warning
If the UP-DF550 should be used adjacent to or stacked with
other equipment, it should be observed to verify normal
operation in the configuration in which it will be used.
4
Guidance and manufacturer's declaration-electromagnetic emissions
Compliance
Group 1
Class B
Class A
Complies
Warning
The use of accessories and cables other than those
specified, with the exception of replacement parts sold by
Sony Corporation, may result in increased emissions or
decreased immunity of the UP-DF550.
Electromagnetic environment-guidance
The UP-DF550 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
The UP-DF550 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.