placed onto an insulated surface. Otherwise, leakage currents to earth might influence the
measuring result.
8.2.2
Testing of devices of protection class II (shock-proof) and of devices of protec-
tion class III (safety extra-low voltage)
-
Testing of devices without protective conductor and with accessible conductive parts
-
Connect the test object to the test socket 1 of the BENNING ST 720.
-
Establish a connection between the 4 mm test socket 9 and a metal part of the test object
by means of the test cable with alligator clip.
-
For mains operating mode (contact current by means of the direct measurement method,
test object in operation!): Connect the plug of the mains connection cable with the socket K
and the shock-proof plug with a protected shock-proof socket (230 V, 50 Hz, 16 A).
-
If necessary, the testing voltage of the R
reduced to 250 V
on the display 8 . Press the key again to switch back to the preset testing voltage of 500 V
-
Switch the test object on.
-
Press the
key 3 to start the automatic testing procedure.
-
If "Lo LOAD" is shown on the display, please check whether the test object is switched on.
-
Press the key 3 to continue the testing procedure in case of the load being too low (R
> 6 kΩ).
-
If "HIGH LOAD" is shown on the display, this indicates an excessive load (R
I
(I
) > 16 A) of the test object. There might be danger of a short-circuit or of an earth
LAST
LOAD
fault. Check whether there is a short-circuit between the external conductor (L) and neutral
conductor (N) of the test object.
-
If there is no short-circuit, you can continue with the testing procedure by pressing the
-key 3 .
-
If the insulating resistance R
next to the R
ISO
BENNING ST 720 in mains operating mode:
-
The BENNING ST 720 interrupts the testing procedure after the R
measurement and requests the user to switch the mains voltage of 230 V to the test socket
1 by showing a flashing "I
press the
-key 4 to measure the contact current I
-
The contact current by means of the direct measurement method only starts as soon as the
mains voltage is correct applied.
step 1 of 2:
-
After a measuring time of 5 seconds or by pressing the key 4 , mains polarity will be re-
versed and the contact current will be measured with reversed mains voltage ("L/N" - "N/L").
The highest measured value of both measurements will be displayed. (step 2 of 2)
08/ 2010
by means of the
DC
is higher than the admissible limiting value, a
ISO
symbol.
" symbol. - Make sure that the test sample is protected and
LEAK
BENNING ST 720
(insulating resistance) measurement can be
ISO
-key 5 . The selected testing voltage is briefly shown
(direct measurement method).
LEAK
<< 14 Ω,
L-N
appears
(insulating resistance)
ISO
.
DC
L-N
25