Ottobock 2WR95 Gebrauchsanweisung Seite 10

Vorschau ausblenden Andere Handbücher für 2WR95:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
5.1 Adjusting the adapter
CAUTION
Improper assembly of the screw connections
Fall due to breakage or loosening of the screw connections
► Clean the threads before every installation.
► Apply the specified installation torque values.
► Observe the instructions for securing the screw connections and
the use of the correct length.
CAUTION
Incorrect processing of tube
Fall due to damage to the tube
► Do not clamp the tube into a vice.
► For shortening the tube, use only a tube cutter.
>
Recommended tools and materials:
710D4 torque wrench, 719R3 tube cutter, 718R1 tube deburrer,
degreasing cleaning agent (e.g. 634A3 Acetone)
1) Use the tube cutter to shorten the tube according to the patient's
measurements.
2) Deburr the inside and outside of the cut edge with the tube
deburrer.
5.2 Mounting the adapter
CAUTION
Incorrect mounting of the tube
Fall due to breakage of load-bearing components
► Slide the tube all the way to the stop in the intended prosthetic
component when mounting.
>
Recommended tools and materials:
710D4 Torque Wrench, degreasing cleaner (e.g. 634A3 Acetone)
1) Loosen the tube clamp cap screw by 2 turns.
10 | Ottobock
2) For final assembly: clean the insertion range of the tube clamp
adapter and tube adapter with degreasing cleaner.
3) Fully insert the tube adapter into the tube clamp adapter up to the
stop. Align the clamping slot:
Tube clamp adapter: anterior
Tube clamp adapter, sliding: anterior or medial
4) Tighten the tube clamp cap screw (installation torque: M4: 5 Nm,
M5: 10 Nm, M6: 13 Nm).
5.3 Installation in the modular prosthesis
CAUTION
Improper assembly of the screw connections
Fall due to breakage or loosening of the screw connections
► Clean the threads before every installation.
► Apply the specified installation torque values.
► Observe the instructions for securing the screw connections and
the use of the correct length.
>
Recommended tools and materials:
710D4 torque wrench, 636K13 Loctite®
► Align the pyramid receiver of the tube adapter distally in the pros­
thesis. Align the clamping slot of one tube clamp adapter anteri­
orly.
The pyramid adapter is fixed with the set screws of the pyramid receiv­
er.
Required tools and materials:
>
710D4 torque wrench, 636K13 Loctite®
1) Trial fitting:
Screw in the set screws.
Use the torque wrench to tighten the set screws (10 Nm).