Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration - electromagnetic immunity
The [KD-735] is intended for use in the electromagnetic environment
specified below. The customer or the user of the [KD-735] should assure
that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U T is the a.c. mains voltage prior to application of the test level.
ENGLISH
IEC 60601
Compliance
test level
level
± 6 kV contact
± 6 kV
± 8 kV air
contact
± 8 kV air
3 A/m
3 A/m
Electromagnetic environment -
guidance
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment