7-3 Lubrication
NSK PANA SPRAY Plus
Pana-Max2 QD/Pana-Max2 PTL/Pana-Max2 KV
Apply NSK PANA SPRAY Plus every time after each use and/or before
autoclaving.
1) Remove the bur from the handpiece.
2) Insert the Spray Nozzle into the Spray Port nozzle on the can.
3) Insert the Spray Nozzle in rear of the handpiece. Hold the
handpiece and spray for approximately 2-3 seconds. Apply
lubricant until it expels from the handpiece head for at least 2
seconds. (Fig.11)
Pana-Max2 M4/Pana-Max2 B2
Apply NSK PANA SPRAY Plus every time after each use and/or before
autoclaving.
1) Remove the handpiece from the hose.
2) Mount the arrow-head spray nozzle tip into the spray can port.
(Fig.12)
3) Insert the Spray Nozzle into the Drive Air Port of the handpiece.
Hold the handpiece and spray for approximately 2-3 seconds.
6
Apply lubricant until it expels from the handpiece head for at least
2 seconds. (Fig.13)
CAUTION
When applying spray be sure to hold the handpiece firmly to
prevent the handpiece from slipping out of the hand due to the
spray pressure.
Hold the spray can upright.
Lubrication Oil (Pana-Max2 M4/Pana-Max2 B2)
1) Remove the handpiece from the hose.
2) Apply 1-2 drops of lubrication oil into the Drive Air Port shown in
the Fig.13.
3) Connect the handpiece to the hose. Operate the handpiece for
about 10 seconds.